Module 08: Quadrupole Mass Filter Simulator

← Academy
Quadrupole Controls
Four rods, two voltages, one survivor. The quadrupole mass filter uses a combination of DC and RF voltages to create a mass-dependent stability window. Only ions of the selected mass oscillate stably and reach the detector. Everything else crashes into the rods.
28 amu
R = 50
RF Voltage (V)
--
DC Voltage (U)
--
U/V Ratio
--
Rod Radius r0
6.35 mm
RF Frequency
2.0 MHz
Resolution M/dM
--
Stability Diagram
Ion Trajectory (cross-section)
Drag the sliders above to explore how the quadrupole selects ions by mass.
Mass Spectrum Builder
Hit SCAN to sweep the RF voltage from low to high, building a mass spectrum bar by bar -- exactly like your MKS MV2 does on every bake station.
Sources · 3
  1. [A]P.H. Dawson, Quadrupole Mass Spectrometry and Its Applications, Elsevier, 1976 (reprinted AIP, 1995).
  2. [A]NIST Chemistry WebBook, Mass Spectra.
  3. [B]MKS Instruments, MicroVision 2 Residual Gas Analyzer : Operating Manual, 2019.